Thickness Quantitative Flash X-ray Imaging of Molten Metal Fragmentation
FXR reveals droplet breakup mechanisms to help prevent hazardous events
ABSTRACT: Flash X ray (FXR) imaging enables visualization of fast, opaque processes such as molten metal fragmentation in water. By applying proper calibration, FXR can be used not only for imaging but also for quantitative measurement of material thickness and mass distribution.
This application note describes the calibration procedure required to convert X ray intensity into material thickness, using molten tin–water interaction experiments as an example. The approach demonstrates how accurate calibration and normalization transform radiographs into time-resolved quantitative data.
Courtesy:
This application note was made in collaboration with Professor David L. Frost (david.frost@mcgill.ca) at the Department of Mechanical Engineering, McGill University, Montreal, QC, Canada.
Thickness calibration for converting radiographs into mass distribution
Scandiflash core equipment
Additional equipment
Kodak XAR-5 high-speed X-ray film
Kodak Mini-R intensifying screen and cassette
High-speed camera
Thickness Calibration Quick Checklist
- Choose correct FXR system and setup for appropriate thickness range
- Fix setup (geometry, shielding, materials)
- Acquire calibration image (including whole image + step wedge)
- Extract intensity vs thickness
- Fit calibration curve
- Normalize images (background + flat-field)
- Apply to experiment
- Respect limits
Droplet breakup is far more than a visual curiosity, it is a key multiphase flow phenomenon that governs how fluids mix, exchange heat, and form new interfaces. As droplets fragment, they create surface area that accelerates thermal and chemical interactions, influencing processes from industrial cooling to melt–coolant interactions where vapor explosions may occur.
Flash X ray radiography enables the study of these systems by penetrating vapor bubbles and dense fragment clouds, revealing internal structures that cannot be observed optically. Beyond visualization, it allows quantitative analysis by converting image intensity into material thickness, and thereby estimating mass distribution within the evolving droplet.
Choice of system and setup
The measurable thickness range is determined by the selected FXR system and operating voltage (kVp), which defines the energy spectrum and penetration capability. Selecting appropriate conditions ensures that the relevant fragment structures fall within the measurable range.
Measurement setup
The full experimental configuration must be fixed prior to calibration, including geometry, shielding, and all materials in the beam path. In this work, the X ray head was positioned 50 cm in front of the droplet and the detector 6 cm behind it, providing a balance between contrast and penetration.
Calibration image acquisition
A reference image is recorded using a calibration target with known thickness steps placed within the field of view. This image contains both the object and the thickness reference under identical conditions.
Intensity–thickness relation
Since flash X ray sources produce a broad spectrum, the measured intensity is given by:
I = ∫S(E)e–µ(E)x dE
Rather than solving this analytically, intensity values corresponding to known thickness steps are extracted from the calibration image.
Calibration curve
These data are used to generate an empirical mapping between normalized image intensity and material thickness:
x = f(Inorm)
Normalization
All images are corrected using a combined normalization step:
Inorm = (Imeas – Ibg) ⁄ (I0 – Ibg)
This removes background signal and compensates for spatial variations in beam intensity.
Application to experiment
The calibration function is applied pixel wise to radiographs of the fragmentation process, converting greyscale intensity into line of sight thickness and enabling spatial mapping of the droplet mass.
Using this approach, the thickness analysis reveals how mass is redistributed during fragmentation. Initially, material is stripped from the droplet surface, forming a fragment cloud. At low coolant velocities, vapor bubble collapse disperses fragments into a more uniform distribution, while at intermediate velocities mass remains concentrated near the droplet core. At high velocities, vapor formation is suppressed and fragmentation is dominated by hydrodynamic stripping, producing a more continuous downstream distribution.
References:
Cunningham MH and Frost DL (2023), Fragmentation of a molten metal droplet in an ambient water flow. Front. Phys. 11:1171267.
doi: 10.3389/fphy.2023.1171267
Cunningham MH. Effect of coolant velocity on the fragmentation of single melt drops in water. MEng. Thesis. Montreal: McGill University (1996).
We're here to help you
Choosing Scandiflash means you get a flash X-ray system that’s built to meet your needs both today and tomorrow as well as a team that’s ready to support you.
Want to know more about taking your research to the next level? We’re curious to hear what you can imagine doing and seeing with our technology, so get in touch with us through our website or just give us a call at +46 18 557510.
Researching dynamics?
Sometimes you only get one chance – make it count.About Scandiflash
For over 50 years, Scandiflash has been pioneering flash X-ray technology to help scientists and researchers around the globe to see the nearly impossible. Scandiflash Flash X-Ray Systems generate extremely short pulses used to capture dynamics in the harshest of conditions, ranging from indoor lab setups to outdoor large-scale firing ranges. The systems are modular and can be tailored to meet your requirements for number of pulses and peak energy conditions.